The Trace101 is a hydrogen tracer gas detector, approved for ATEX Div. 1 or zone 0 operation. The portable Trace101 Leak Detector makes the revolutionary Hydrogen Method available
for leak testing in the toughest of environments, including explosion hazardous areas. Finding leaks down to bacteria size, Trace101 is the ultimate tool for a wide range of demanding applications. The intrinsically safe Trace101 leak Detector brings accuracy, flexibility and simplicity into potentially explosive environments. It meets the ATEX directive for use in Div 1 or zone 0-ATEX Classification Ex ia, IIC T4. Versatile and easy to operate, the Trace101 is designed to withstand rough handling in the field and is easily carried. It has a chargeable Ni-MH battery providing up to four hours of operation. The new ergonomically designed hand probe allows for fast homing in on suspected leak areas and exact leak locating and quantification. The recommended tracer gas is a low cost standard mixture of 5% hydrogen (H2) and 95% nitrogen (N2), available from major gas suppliers. This mixture is nonflammable, non-toxic and environmentally friendly. Simple solution for demanding applications Trace101 is the ultimate choice for a wide range of industrial leak detection applications. For the first time ever, the industry has got a safe, accurate and yet simple solution for leak testing of the complete industry, the petrochemical area, hydrogen production plants and on oil rigs. The Trace101 is the perfect tool for locating leaks on all types of valves, joints and other components. Leak testing of hydrogen cooled generators and fuel cells are typical production and maintenance applications.

Features
Trace101 – Specifications
Detection Performance
Lowest detectable hydrogen concentration 1 ppm (parts per million)
Lowest detectable leak rate 1 × 10-6 std cc/sec
Standard gas mixture 5% H2 and 95% N2
Timing
Response time 1 second
Warm up time (including self-check) 1 minute (approx.)
Power
Battery unit Sealed rechargeable battery
Battery runtime 4 hours
Battery charger 115–230 Vac, 50–60 Hz / 12 Vdc
Battery safety note Battery must not be charged in explosive area
Calibration
Calibration & alarm setpoint Field calibration, verified to NIST standard calibrated leak
Operating Conditions
Temperature 0°C (32°F) to +50°C (122°F)
Relative humidity 90% maximum
Physical Characteristics
Weight (with battery) 600 g (1.32 lbs)
Size 235 × 118 × 45 mm (9.2″ × 4.6″ × 1.8″)
Certification
CSA Ex i – Class 1 Div 1 Group C–T4
Vmax = 7.5 V
Installation drawing DP 100 01 02 01 01 00
Manufacturer SunAero
Europe (ATEX) INERIS 08 ATEX 0039 Ex
II 1 G Ex ia IIC T4 Ga
Tamb: 0°C to +50°C, Zone 0, Category 1
Standards NF EN 60079-0:2009
NF EN 60079-11:2012